VERITY 1810 Conductivity Monitor

The VERITY® 1810 Conductivity and pH Monitor is a state-of-the-art detection instrument for biological purifications with a large dynamic range and fast response time for real-time gradient monitoring. The monitor is a compact instrument with an integrated conductivity sensor, an external connection for an optional pH sensor, and analog output connections for conductivity and pH. The monitor displays crucial parameters for your experiment in real-time, allowing you to protect your valuable samples, ensure their purity, and optimize each step of your method from beginning to end.

Efficiently Capture Critical Data

Monitor column equilibration, ensure gradient formation, and control the quality of your eluent. A single plug-and-play device displays both conductivity and pH during the run.

Integrate Into Any System

Perform oligonucleotide, monoclonal antibody (mAb), and other biological separations via simple analog outputs.

Conserve Bench Space

The VERITY 1810 Monitor is the size of a deep-well microplate.

Specifications

VERITY 1810 Conductivity Monitor

Analog Outputs Two 0–1V dedicated analog outputs: one for streaming conductivity data and the other for streaming pH data
Communication USB
Conductivity Flow Cell Accuracy ± 4% or ± 20 μS/cm,whichever is greater,within 1 μS/cm–300 mS/cm at 25°C
Pressure psi bar MPa
200 13.79 1.38
Readable Range lμS/cm - 999.99mS/cm
Specification Range lμS/cm - 300 mS/cm
Wetted Materials PEEK, Titanium
Volume 54.1 μL
Data Acquisition 2, 5, or 10 points/sec with TRILUTION® LC Software
Dimensions (W x D x H) 13.4 x 10.3 x 7.1 cm (5.3 x 4.1 x 2.8 in.)
Environmental Conditions Indoor use
Altitude: up to 2000 m
Temperature range: 4°C–40°C
Humidity: Maximum relative humidity 80% for temperatures up to 31°C, decreasing linearly to 50% relative humidity at 40°C
Front Panel Three line display
Monitor Channels Three: Conductivity, pH, and Temperature
pH Flow Cell Accuracy ±0.1 units at 25°C
Maximum Drift 0.1/10 hours at constant conditions
Pressure psi bar MPa
100 6.89 0.69
Readable Range pH 0-14
Specification Range pH 2-12
Wetted Materials Acrylic, PVDF, pH Glass, epoxy, BUNA-N
Volume 49.2μL
Power Requirements USB powered instrument
Instrument Power Requirement: 5V DC (1W)
Recommended Flow Rate Range 0.5-150 mL/min
Safety and Compliance The instrument has been certified to safety standards specified for Canada, Europe, and the United States. Refer to the instrument rear panel label and the Declaration of Conformity document for the current standards to which the instrument has been found compliant.
Software Control PC control via USB and TRILUTION® LC Software
Weight 0.34 kg (0.75 lbs.)

Supporting Literature

User Guides
VERITY 1810 User's Guide Download (EN)    

2023 Gilson Product Guide

The 2023 Gilson Product Guide is available now. Discover best sellers for your Liquid Handling, Purification, and Extraction needs!

Download

Gilson Services and Calibration Programs

Gilson is committed to quality and complete customer satisfaction with all of our products and services. Get system support, training programs and pipette service from the people who know Gilson products the best. We are here to help and provide you with dependable solutions to accommodate your needs.

LEARN MORE

COVID-19 Testing

COVID-19 Testing: Liquid Handling Solutions to Support Your Efforts






LEARN MORE
The Gilson website uses cookies, as explained in our Personal Information Protection Policy. If you agree to our use of cookies, please close this message and continue to use the site - ACCEPT & CLOSE
© 2024 Gilson, All Rights Reserved. Privacy Policy | Terms of Service | Terms of Use